The ASME B89.4.19 System Calibration (960-02589) is an additional calibration standard FARO uses in conjunction with the FARO® Standard Calibration (ACC-00). The Laser Tracker must go through the FARO Standard Calibration before it can go through the ASME B89.4.19 System Calibration (960-02589). This article contains information about both standards.
All FARO factory service centers are ISO/IEC accredited. The scope of our accreditation does not include the FARO Standard Calibration (ACC-00) process, therefore our accreditation information is not listed on the certificate. The optional ASME B89.4.19 system calibration is accredited. Click here for in depth information about our ISO accreditation.
To get a copy of the ASME B89.4.19 Standard, download it from American Society of Mechanical Engineers web site.
This standard provides a consistent and accepted method of qualifying the accuracy of Laser Tracker measurements to National Measurement Institutes Standards (such as NIST, NPL, DKD, Spring Singapore). The standard specifies using two traceable artifacts: traceable scale bars at predetermined distances and locations, as well as National Measurement Institutes Standard traceable interferometers.
Once this calibration is completed, the Tracker has National Measurement Institutes Traceable measurements instead of National Measurement Institutes Traceable components.